BlueBlink specifications

Specifications

The table below shows the main technical specifications of the EBERS BlueBlink oxygen sensors.

 

BlueBlink oxygen sensors
Optical reader
Sampling time 1 – 10min (in 1 min. increments).
Wavelength ilumination LEDs 450nm
Sensing spots
Analyte Gaseous & disolved oxygen
Sterility Pre-sterilized
Measurement range 0-21% O2
0-230 hPa
16 mg/L

Resolution at:

  • 1 % O2
  • 10 hPa

 

  • 0.01 %
  • 0.1 hPa
 Resolution at:
  • 20.9 % O2
  • 210 hPa
 
  • 0.03 %
  • 0.3 hPa
Accuracy 2.5 %
Measurement T range 0 – 50 ºC
Calibration Two-point calibration in oxygen-free  environment (nitrogen, sodium sulfite) and air-saturated environment
Number of measurements ~10.000
Attachment

Self-adhesive, spots attach well to most cell culture surfaces.
Non self-adhesive spots & sterile glue are also available.
Disposable spots.

Incompatibilities Organic solvents, chlorine gas
No cross-sensitivity with: pH, CO2, H2S, SO2, ionic species
Operation conditions
Temperature range 0 – 50ºC
Humidity range 0 – 100%
Pressure range 0 – 2000hPa
Storage condiotions
Temperature Room temperature
Software requirements
Windows 7 or higher
Dimensions
Optical reader 24 x 24 x 36 mm
Sensing spot 15 mm in diameter

Please contact us for further information if you want to know more about the BlueBlink sensors.