The table below shows the main technical specifications of the EBERS BlueBlink oxygen sensors.
BlueBlink oxygen sensors | |
---|---|
Optical reader | |
Sampling time | 1 – 10min (in 1 min. increments). |
Wavelength ilumination LEDs | 450nm |
Sensing spots | |
Analyte | Gaseous & disolved oxygen |
Sterility | Pre-sterilized |
Measurement range | 0-21% O2 0-230 hPa 16 mg/L |
Resolution at:
|
|
Resolution at:
|
|
Accuracy | 2.5 % |
Measurement T range | 0 – 50 ºC |
Calibration | Two-point calibration in oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment |
Number of measurements | ~10.000 |
Attachment |
Self-adhesive, spots attach well to most cell culture surfaces. |
Incompatibilities | Organic solvents, chlorine gas |
No cross-sensitivity with: | pH, CO2, H2S, SO2, ionic species |
Operation conditions | |
Temperature range | 0 – 50ºC |
Humidity range | 0 – 100% |
Pressure range | 0 – 2000hPa |
Storage condiotions | |
Temperature | Room temperature |
Software requirements | |
Windows 7 or higher | |
Dimensions | |
Optical reader | 24 x 24 x 36 mm |
Sensing spot | 15 mm in diameter |
Please contact us for further information if you want to know more about the BlueBlink sensors.